Catalog Number 05.001.224
Device Problems Corroded (1131); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During evaluation it was determined that the initial reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, the malfunctions found during service and evaluation have been confirmed.The assignable root cause was determined to be due to component wear.Udi:(b)(4).
 
Event Description
It was reported from germany that during service and evaluation, it was determined that the saw attachment device was frozen/will not move.It was further determined that the device failed pretest for general condition, check for mechanical free movement, check general function in the running mode and check the oscillation frequency with the frequency meter.It was noted in the service order that the device had corrosion/rusting/pitting.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAGITTAL SAW ATTACHMENT, LONG, FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key17512427
MDR Text Key320917319
Report Number8030965-2023-10182
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819377929
UDI-Public07611819377929
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2023
Date Manufacturer Received08/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/26/2013
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-