SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
|
Back to Search Results |
|
Catalog Number P310NUS |
Device Problems
Break (1069); Key or Button Unresponsive/not Working (4063)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the device's knobs are broken on the unit.Patient involvement is unknown.
|
|
Manufacturer Narrative
|
Health effects codes, updated.
|
|
Event Description
|
Additional information received via email.Event date remains unknown.There was no patient injury.No further details provided.
|
|
Manufacturer Narrative
|
Other, other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device was received in with all small knobs cracked.The peep and cpap control knobs were broken off.The complaint was confirmed.The root cause was user interface from impact to the device.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced small knobs and caps.Replaced variable relief valve assembly, front panel, o-rings, input filter and mri battery.Performed preventative maintenance (pm), calibrated, and cleaned the device.The device passed all functional and delivery tests.
|
|