Catalog Number P310NUS
Device Problems Break (1069); Key or Button Unresponsive/not Working (4063)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Other, other text: b3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the device's knobs are broken on the unit.Patient involvement is unknown.
 
Manufacturer Narrative
Health effects codes, updated.
 
Event Description
Additional information received via email.Event date remains unknown.There was no patient injury.No further details provided.
 
Manufacturer Narrative
Other, other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(6).One device was received in with all small knobs cracked.The peep and cpap control knobs were broken off.The complaint was confirmed.The root cause was user interface from impact to the device.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced small knobs and caps.Replaced variable relief valve assembly, front panel, o-rings, input filter and mri battery.Performed preventative maintenance (pm), calibrated, and cleaned the device.The device passed all functional and delivery tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC PLUS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17512472
MDR Text Key320919672
Report Number3012307300-2023-08050
Device Sequence Number1
Product Code BTL
UDI-Device Identifier35019315107263
UDI-Public35019315107263
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberP310NUS
Was Device Available for Evaluation? No
Date Returned to Manufacturer07/27/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-