It was reported that this right atrial lead oversensed noise leading to episodes of atrial tachycardia response.The noise appeared consistent with electromechanical interference (eml).It was noted that the patient had a glucose monitor and an insulin pump.A technical services representative discussed that the noise could potentially be a result of oversensing of the respiratory rate trend (rrt) signal.Ts discussed emi considerations and suggested that, if the noise continued, rrt could be programmed off.Ts also noted that impedance was within normal limits.No adverse patient effects were reported.Evidence suggests this lead remains in service.This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|