Product event summary: the afapro28 balloon catheter with lot number 06867 was returned and analyzed.Visual inspection of the balloon catheter was performed, and no anomalies were identified during external visual inspection of the balloon, shaft, and handle segments.The balloon catheter smart chip data was reviewed.Data indicated the balloon catheter was used for nine applications.However, the balloon catheter usage date recorded on the smart chip (b)(6) 2023 did not correspond to the event date.The balloon catheter was recognized and passed the electrical integrity verifications and performance test.The balloon catheter completed the inflation, ablation, and thawing phases with no console system notices generated.No performance issues were identified.All pressure and flow were in range and temperature curve had no oscillation or overshoot.During inspection of the shaft segment, a guide wire lumen kink/twist was observed 1.1614 inches proximal to the catheter tip.Pressure testing did not identify any leakage from the kink.In conclusion, the reported kink was confirmed and the balloon catheter failed the returned product inspection due to a kink/twist observed on the guide wire lumen.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|