This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the raised forceps elevator wires due to a cut/fray was that the k-wire had accumulated fatigue and was broken by repeated manipulation of the forceps elevator.The event can be detected/prevented by following the instructions for use which state: ¿3.2 inspection of the endoscope: inspection of the forceps elevator mechanism: inspection for smooth operation: visually confirm that the portion of the elevator wire extending from the distal end of the endoscope is not broken, frayed, or bent (see figure 3.5).If any damage (broken, frayed, or bent portion) is observed on the elevator wire as shown in figure 3.5, do not use the endoscope.¿ olympus will continue to monitor field performance for this device.
|