Model Number JF-260V
Device Problem Material Frayed (1262)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned for evaluation and the customers allegation was confirmed.It was noted that there was waterproof failure due to a cut on the bending section rubber and there was a white scratch on the image due to a broken charge coupled device unit.The light guide lens was broken and the electrical connector was corroded.The forceps raising angle was out of standard due to loose forceps elevator wire.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported to olympus field service engineer, that the evis lucera duodenovideoscope had raised forceps elevator wires due to a cut/fray.The issue was found during reprocessing.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the raised forceps elevator wires due to a cut/fray was that the k-wire had accumulated fatigue and was broken by repeated manipulation of the forceps elevator.The event can be detected/prevented by following the instructions for use which state: ¿3.2 inspection of the endoscope: inspection of the forceps elevator mechanism: inspection for smooth operation: visually confirm that the portion of the elevator wire extending from the distal end of the endoscope is not broken, frayed, or bent (see figure 3.5).If any damage (broken, frayed, or bent portion) is observed on the elevator wire as shown in figure 3.5, do not use the endoscope.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17553693
MDR Text Key321725941
Report Number9610595-2023-11722
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberJF-260V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse