The initial reporter received a questionable elecsys ft4 iv assay result from one patient sample tested on the cobas e 801 module.The initial result was reported to the physician who deemed the result implausible, prompting the rerun of the patient sample.The patient sample was rerun on the module and two other modules (module 1 and module 2).The initial result from the module, with the patient sample in a hitachi cup on a tube, was 0.129 ng/dl.The first repeat result from the module, with the patient sample in a hitachi cup on a tube was 5.21 ng/dl.The second repeat result from module 1 using a different reagent lot and with the patient sample in a microcup was a data flag with no numeric result.The third repeat result from module 1 using a different reagent lot and with the patient sample in a microcup was 5.67 ng/dl.The fourth repeat result from module 2, with the patient sample in a micro cup was 5.88 ng/dl.The third and fourth repeat results confirmed the first repeat result.
|