FUJIFILM HEALTHCARE CORPORATION OASIS MRI SYSTEM; SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Model Number N/A
Device Problem Excessive Heating (4030)
Patient Problem Superficial (First Degree) Burn (2685)
Event Date 07/18/2023
Event Type  Injury  
Event Description
On july 26, 2023, fujifilm healthcare corporation became aware of an event involving oasis mri system.It was reported that the patient complained of a burning sensation and first-degree burns around her neck where the coil surrounded the area following a c-spine exam.There was no death associated with the event.
 
Manufacturer Narrative
The investigation concluded that there are no problems with the device or components, and all functionality was within normal specifications.The reported issue is not an isolated occurrence, and fujifilm is monitoring the facility for similar issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OASIS MRI SYSTEM
Type of Device
SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
FUJIFILM HEALTHCARE CORPORATION
2-1 shintoyofuta
kashiwa-shi, chiba-ken 277-0 804
JA  277-0804
Manufacturer (Section G)
FUJIFILM HEALTHCARE CORPORATION
2-1 shintoyofuta
kashiwa-shi, chiba-ken 277-0 804
JA   277-0804
Manufacturer Contact
yasuo okamoto
2-1 shintoyofuta
kashiwa-shi, chiba-ken 277-0-804
JA   277-0804
MDR Report Key17560810
MDR Text Key321309923
Report Number3018423337-2023-00008
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211406
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Distributor
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 08/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age48 YR
Patient SexFemale
Patient Weight118 KG
-
-