The visual and functional check of the handpiece prior to the repair showed that the product was running out of specification.It was found that the ball bearings have been worn out, hence they have been grinding and vibrating.The worn-out bearings caused higher friction and hence heat up of the head.The visual inspection showed also that the head had a dent which causes at least misalignment of the inner components and stronger wear.It is a sign that the product received a strong external force, e.G.By dropping it.To avoid such issues the user instruction contains already several notes, warnings and requests how to prepare the handpiece for each treatment and how to use it: warning: hazards for the care provider and the patient.In the case of damage, irregular running noise, excessive vibration, untypical warming or when the cutter or grinder cannot be held.-> do not use further and notify service.Caution: burning hazard from hot instrument head or hot instruments cover.If the instrument overheats, burns may arise in the oral area.->never contact soft tissue with the instrument head or instrument cover the following guidelines must be observed to ensure save use of the electrically driven contra-angle handpieces: >the service instructions for contra-angle handpieces must be precisely following when using kavo spray or quattrocare care systems.> before each use, the contra-angle handpiece must be checked for external damage.> before each use, perform a test run with the contra-angle handpiece, and watch for atypical heating and unusual noise and vibration.> immediately stop using contra-angle handpieces that act unusual.> never press the pushbutton during operation.This also includes lifting the cheek or tongue! to ensure proper function, the medical device must be set up according to the reprocessing methods described in the kavo instructions for use, and the care products and care systems described therein must be used.Kavo recommends specifying a service interval at the dental office for a licensed shop to clean, service and check the functioning of the medical device.This service interval depends on the frequency of use and should be adjusted accordingly.
|
When handpiece was sent in for repair the customer letter stated that it was getting hot, burning patient.During contact with the dental office dentist informed that during a crown preparation on tooth #14 the patient received a burn on upper lip, left side inside oral cavity.The burn was only small in size, and the patient was told to use some otc neosporin to avoid inflammation.
|