(b)(4).Date sent: 8/17/2023.D4 batch #: x95f6z.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the tissue pad detached and not returned but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.Additionally, the device was returned with the blade tip damaged.The device was connected to a test handpiece and a gen11, and the device did activate during functional testing.The device opens or closes as expected.The device was disassembled to verify the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause for this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in the removal of the pad during use.If the device is activated across a clip, staple line, or other metal in the jaws, the blade could get damaged, subsequent activations may increase the severity of the blade damage.Once minor blade damage has occurred can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These screen alerts can be such as, ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch x95f6z, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|