Model Number ALLURA XPER FD20 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2023 |
Event Type
malfunction
|
Event Description
|
It was reported to philips that the plastic cover on the arm of operating room light fell into the sterile field.The system was in clinical use.There was no reported patient or user harm.A philips field service engineer (fse) visited the site and re-attached the covers to the light arm.The device was returned to expected functionality.
|
|
Manufacturer Narrative
|
Philips has investigated this complaint.According to the additional information added the system was not in clinical use.The philips field service engineer (fse) inspected the system onsite and confirmed that a plastic cover on the arm of or lite fell off and into the sterile field during case.Upon some functional test fse found two screws which hold the cover came loose due to which the cover fell.Fse fixed cover back into the light room after it fell off by tightening the screws, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
|
|