SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported the image of the evis exera ii video system center had interruptions or image contacts.There was no reported patient harm or impact due to this event.
 
Manufacturer Narrative
Attempts to retrieve additional information from the customer are in progress.During device evaluation at olympus, it was found the complaint was not confirmed.The device was tested and no faults were found; however, there was a lot of dust inside the device that could cause overheating and picture issues.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to fields h4, h6.Corrected field: e1 (phone number).A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 17 years since the subject device was manufactured.According to the facts found out from the investigation, the reported issue was not reproduced during tests, and the device passed all tests.Therefore, investigators were unable to identify the root cause.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17570927
MDR Text Key321417825
Report Number3002808148-2023-08582
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-