Catalog Number 04.043.245S
Device Problems Difficult to Advance (2920); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2023, they were inserting the nail and the poly inside the proximal part of the nail seemed to be damaged.They tried inserting the ball tip wire through and it would not advance.They removed the nail and noticed the poly was damaged.There was 3 minute delay and got a new nail off the shelf and placed it successfully.No fragments generated and case was completed successfully.This report is for one (1) tibial nail-advanced / 10 375 / sterile.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional product code: hwc.D9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d4, h4 a manufacturing record evaluation was performed for the finished device product code: 04.043.245s-us lot number: 3297p13 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 15/12/2022 manufacturing site: jabil bettlach expiry date: 30/11/2032.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from the photo.Visual analysis of the photo revealed that the tibial nail-advanced / 10 375 / sterile had the proximal poly inlay damaged and deformed.Base on the damaged of the inlay, it can be assumed that the observed condition prevents its mating instrument to be fully inserted as intended.No other defects can be observed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for tibial nail-advanced / 10 375 / sterile.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL-ADVANCED / 10 375 / STERILE
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17572350
MDR Text Key321435144
Report Number8030965-2023-10421
Device Sequence Number1
Product Code JDS
UDI-Device Identifier10886982297808
UDI-Public(01)10886982297808
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201336
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.043.245S
Device Lot Number3297P13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-