Zoll medical corporation evaluated the electrode pads and the customer's report was not replicated or confirmed.Visual inspection of the electrode pads observed some slight discoloration around part of the outer edges of the tin of the front pad.There were no signs of any burns or sparking.The pads passed continuity testing.The pads were scrapped.No clinical data was provided for review.During defibrillation therapy, sparks could occur due to a high measured patient impedance.This could be due to many reasons including, but not limited to, how the skin is prepared, how the electrodes are applied, or poor coupling.Electrode labeling provides instructions for proper electrode application technique.Poor adherence and /or air under the electrodes can lead to the possibility of arcing.It is likely that there is poor coupling between the pads and the patient, but without the device log or additional information regarding patient prep or pads application, root cause cannot be determined.No trend is associated with reports of this type.
|