SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX EXPANDABLE BREATHING CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C45041318J |
Device Problems
Break (1069); Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/28/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during pre-testing, the device had a cracked/broken swivel part.No adverse effects have been reported.
|
|
Manufacturer Narrative
|
Other text: d4: udi section, expiration date and h4: manufacture date are unknown, no information has been provided to date.G5: 510k is blank, device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Device evaluation: three photos were received for investigation, as well as samples without original packaging.The visual inspection found a broken connector.The leaking issue was detected in the leak test.The complaint was confirmed as the leaking air was caused by the broken connector.Based on the analysis conducted of the sample provided, the failure mode could be caused due to improper handling of the product outside of the manufacturing facility.No corrective actions were required.A dhr review was unable to be completed as the lot number was not provided.
|
|