Catalog Number 530.710 |
Device Problems
Mechanical Problem (1384); Intermittent Loss of Power (4016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of saw blade locking knob not working properly was not confirmed.Therefore, an assignable root cause was not determined.However, the device experiencing intermittent operation, identified during service and evaluation was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.Udi: (b)(4).
|
|
Event Description
|
It was reported that the saw blade locking knob of the battery oscillator device was not working properly.During in-house engineering evaluation, it was observed that the device experienced intermittent operation.It was further determined that the device had corrosion/rusting/pitting, substance/debris/cleaning/sterilization, would not run, and could not secure/lock the cutter.It was further determined that the device failed pretest for check for correct clamping nut, check the quick coupling for saw blades, check function of device and check oscillation frequency with frequency meter.It was not reported if the device was used in surgery, or if there was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|