DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Failure to Power Up (1476); Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that outside of surgery the cart would not power up, when plugged into wall outlet.Found the power inlet module had failed.Due diligence is complete.No additional information is available.At product evaluation investigation some slight charring was found on the fuse compartment.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).The investigation is complete.This is an initial final report submission.Review of the most recent repair record determined unit would not power on, some slight charring on the fuse compartment.Power inlet module failed and was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|