Catalog Number D133605IL
Device Problems Crack (1135); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2023
Event Type  malfunction  
Manufacturer Narrative
The bwi product analysis lab received a video for evaluation.Device investigation details: a video was received for evaluation following biosense webster's procedures.According to video provided by the customer, the pebax was observed with broken with a piece sticking out.A manufacturing record evaluation was performed for the finished device 31001842m number, and no internal action related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch¿ electrophysiology catheter in which device damage occurred.Before the operation, the tip of the device had been found damaged when the package was opened--the outer box/packaging was not damaged.It seemed like the outer layer of the device was broken and detached.The catheter pebax itself was not damaged.A second device was used to complete the operation.There was no adverse event reported on patient.The reported device wasn't used on the patient.The broken/detached component on the catheter tip is mdr-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17592645
MDR Text Key321839581
Report Number2029046-2023-01822
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030031/S053
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/23/2024
Device Catalogue NumberD133605IL
Device Lot Number31001842M
Was Device Available for Evaluation? No
Date Manufacturer Received07/26/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SYNAPTIC S0856332L1/N18710E
-
-