• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/23/2023
Event Type  malfunction  
Manufacturer Narrative
Stryker evaluated the customer's device and was able to confirm the reported issue.The qc connector was loose on the therapy pcba connection.The connection was reseated to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.
 
Event Description
The customer contacted stryker to report that their device would not recognize the defibrillation paddles when attempting to use on a patient.In this state the device would not be able to deliver defibrillation therapy if needed.This issue is patient related; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 15 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
dara friedman
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key17593084
MDR Text Key321628144
Report Number0003015876-2023-01497
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873988015
UDI-Public00883873988015
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number15
Device Catalogue Number99577-000047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2023
Date Manufacturer Received07/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-