It was reported that "the contrast agent was found leaking from the part of the catheter near the user's hand during use.As the leak did not stop thereafter, the catheter was removed and replaced with a new one, inserted at the same insertion site to complete the procedure.No harm to the patient was reported.The user was unable to find at which part of the catheter the leak occurred." the patient status is reported as "fine".
|
(b)(4).The reported complaint that the "contrast agent was found leaking from the part of the catheter near the user's hand during use" was confirmed upon inspection of the returned sample.The customer returned a 5fr.80cm berman catheter with the original packaging pouch (inp-3, inp-7) for investigation.The sample was returned in the ups shipping box and was in a sealed ziploc bag (inp-1, inp-2).The sample was loosely packed within the original packaging pouch (inp-5).Upon return, the catheter body was immediately noted ruptured near the junction hub; the body was noted ruptured from approximately 89.1cm to 89.3cm from the distal tip of the catheter (inp-11, inp-12).The supplied control stroke syringe was noted connected to the inflation lumen stopcock; no damage or abnormalities were noted to the returned syringe (inp-8).The inflation lumen stopcock was in the open position (inp-8).The recommended volume capacity of the balloon is 0.75cc (inp-9).Upon microscopic inspection, the balloon appeared typical; no damage or abnormalities were noted to the balloon (inp-10).No condensation was noted within the inflation lumen extension line.No contrast media was noted within the injection lumen extension line.No blood was noted on the interior or the exterior surfaces of the returned sample.No other damage or abnormalities were noted.The inflation lumen was injected with 0.75cc of air using the returned control stroke syringe.The balloon inflated symmetrically (anp-1, anp-2).Both sides of the balloon measured approximately 4mm each.The balloon did meet specifications per graphic of radius ratio less than or equal to 2.0.The inflation lumen was injected with 0.75cc of air using the returned control stroke syringe.The balloon inflated symmetrically (anp-1, anp-2).The balloon deflated in less than 3 seconds when the syringe was removed per specification.No pull away was noted after the tug test.The balloon was placed in water and air was injected into the inflation lumen again.No leak was noted.The ruptured catheter body did not result in damage to the inflation lumen.The catheter's injection lumen was aspirated/flushed, and air was noted leaking from the ruptured catheter body (anp-3).Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the ruptured catheter body.The probable root cause of the complaint was manufacturing related.Capa has been initiated under teleflex's quality system by the manufacturing site to address this complaint issue.Teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a.Corrected data: n/a.
|
It was reported that "the contrast agent was found leaking from the part of the catheter near the user's hand during use.As the leak did not stop thereafter, the catheter was removed and replaced with a new one, inserted at the same insertion site to complete the procedure.No harm to the patient was reported.The user was unable to find at which part of the catheter the leak occurred." the patient status is reported as "fine".
|