• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MERLIN PROGRAMMER; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MERLIN PROGRAMMER; NO MATCH Back to Search Results
Model Number 3650
Device Problem Program or Algorithm Execution Failure (4036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/31/2023
Event Type  malfunction  
Event Description
It was reported, during threshold testing the merlin programmer did not stop the threshold test when the testing button was released, the test continue until minimum output.The patient did not experience any adverse consequences due to the test continuing.The patient was stable.
 
Manufacturer Narrative
The field complaint of programmer functionality issue was not verified as the event could not be reproduced.
 
Manufacturer Narrative
The reported event of the threshold test continuing after release of the test button was confirmed via review of the device session.Based on the information provided, the root cause of the failed decrement capture test was unable to be determined.There were no additional findings related to the event.The top housing of the programmer was replaced and all tests performed were normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17600004
MDR Text Key321714769
Report Number2017865-2023-38992
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-