U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. AZURION; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number AZURION 7 M20
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/24/2023
Event Type  malfunction  
Event Description
It was reported to philips that the plastic cover of the procedure light's arm was loose and needed to be re-attached.The device was in clinical use at the time of the event.No harm to the patient or user was reported.A philips field service engineer (fse) visited the site and reattached the procedure light¿s plastic cover.The device was returned to expected functionality.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the information provided, the issue was observed during a planned treatment.The treatment continued as planned and the issue had no impact on the treatment time.A philips field service engineer (fse) inspected the system onsite and confirmed that the reported problem.Troubleshooting actions determined that the plastic cover for the surgical light in the room was hit by the stand and the cover came off.The fse reattached the plastic cover and tested the movements of the exam light arm.After reattaching the cover, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AZURION
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key17603102
MDR Text Key321735715
Report Number3003768277-2023-04600
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838085268
UDI-Public00884838085268
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172822
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAZURION 7 M20
Device Catalogue Number722079
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-