The inner sheath, for 26 fr.Outer sheath was returned as part of the asset return process.During routine inspection of the device by olympus, the isolation beak was broken.There was no procedural or patient involvement associated with this event.This mdr is being submitted to capture the reportable malfunction found during the device evaluation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the event was confirmed.The damage of the insulation material of the sheath was caused by improper handling, application of excessive force, mechanical impact like fall, shock or similar stress, thermal mechanical overload.Based on the described damage pattern, it can be determined that the insulation tip's damage was caused by mechanical thermal influence (mechanical loading or thermally induced straining).A capa was initiated for the damage at the insulation insert.The instruction manual identifies the following which could have prevented the phenomenon: -4 before use: warning.Infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.-4.1 inspection and testing inspecting the product: visually inspect the product.Make sure that it has: no corrosion.No dents.No scratches.Ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G cracks, fractures).Warning.Risk of injury: impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.Olympus will continue to monitor field performance for this device.
|