E1: initial reporter address, city, phone: (b)(6).Carotid wallstent monorail 8.0-21 was received for analysis.A visual and tactile examination identified a complete separation of the outer sheath of the device located approximately 30mm distal of the main t-valve.This type of damage is consistent with excessive force being applied to the device when attempting deployment.No kinking of the delivery system was identified.The device was returned with the stent in the correct position on the delivery system.The investigator was unable to deploy the stent due to a separation of the outer sheath of the device.
|
Reportable based on device analysis completed on 02aug2023.It was reported that shaft kink occurred.The 90% stenosed target lesion was located in the internal carotid artery.An 8.0-21 carotid monorail was selected for use.However, the delivery shaft was noted to be kinked, and during the procedure, the stent could not be deployed.The procedure was completed with another of same device.There were no patient complications reported.However, returned device analysis found the outer sheath of the device to be separated.
|