(b)(4).Method: the complaint ojr414 optiflow junior 2 cannula was received at fisher & paykel healthcare (f&p) in new zealand for evaluation where it was visually inspected and ftir analysis was carried out.Results: visual inspection of the complaint cannula revealed that one tube was stretched near the swivel grip.A white substance was observed on the outside of this tube.Ftir analysis confirmed that the white material is the adhesive used to attach the cannula tube to the swivel grip.Conclusion: we are unable to determine the cause of the reported event.However, it is most likely caused by the tubing being pulled.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: - appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death.- do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).- ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
|