LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-85 |
Device Problems
Failure to Pump (1502); Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/26/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t, when turned on, displayed an error code associated to patient circuit 2 pump that did not start (power consumption is too low).There was no patient involvement.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in jackson, michigan.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
Through follow-up communication, livanova learned that the error messages occurred after the patient pump 2 replacement following a noise problem.Field service technician was dispatched to the facility and found the replaced patient pump defective.To solve the issue a second new pump has been installed.All functional tests positively passed and the unit was put back into service.Complaints database analysis revealed no similar out of box event over the last 3 years.Based on the above, the root cause of the event was an isolated manufacturing failure of the new patient pump: out of box failure.The issue would in any case be detected during installation and therefore is always detectable before use.The event has been reassessed as not reportable.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See intial report.
|
|