LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problems Failure to Pump (1502); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t, when turned on, displayed an error code associated to patient circuit 2 pump that did not start (power consumption is too low).There was no patient involvement.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in jackson, michigan.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
Through follow-up communication, livanova learned that the error messages occurred after the patient pump 2 replacement following a noise problem.Field service technician was dispatched to the facility and found the replaced patient pump defective.To solve the issue a second new pump has been installed.All functional tests positively passed and the unit was put back into service.Complaints database analysis revealed no similar out of box event over the last 3 years.Based on the above, the root cause of the event was an isolated manufacturing failure of the new patient pump: out of box failure.The issue would in any case be detected during installation and therefore is always detectable before use.The event has been reassessed as not reportable.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See intial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key17610054
MDR Text Key321807286
Report Number9611109-2023-00394
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817900528
UDI-Public(01)04033817900528(11)161115
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-