MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M Back to Search Results
Model Number BI75000021
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/10/2023
Event Type  malfunction  
Manufacturer Narrative
H3,h6) no parts have been received by the manufacturer for evaluation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that they were doing qa testing and image testing with the system and noted that there was a ring artifact that was appearing in the same location of the 3d scans.He did not mention anything about 2d scans being affected. there was no patient involvement.2023-aug-17 interface update details: additional information was received.It was reported that once the fluoro and rad gain calibrations were completed the artifact was no longer present.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key17611482
MDR Text Key321831951
Report Number3004785967-2023-00518
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI75000021
Device Catalogue NumberBI75000021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/10/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-