|
Model Number 45007 |
Device Problem
Suction Problem (2170)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/23/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that aspiration was lost.This 2.4mm jetstream xc catheter was selected for use in a lower extremity angiogram procedure for peripheral artery disease.After approximately four minutes of run time, after three passes in blades down mode, and during the fourth pass in blades up mode halfway through the occlusion, aspiration appeared to be lost as if the catheter was clogged.Device suction could be heard; however, no bodily fluid or saline was being returned.Troubleshooting was performed but was unable to resolve the issue.The procedure was completed with another of the same device.There were no patient complications.
|
|
Event Description
|
It was reported that aspiration was lost.This 2.4mm jetstream xc catheter was selected for use in a lower extremity angiogram procedure for peripheral artery disease.After approximately four minutes of run time, after three passes in blades down mode, and during the fourth pass in blades up mode halfway through the occlusion, aspiration appeared to be lost as if the catheter was clogged.Device suction could be heard; however, no bodily fluid or saline was being returned.Troubleshooting was performed but was unable to resolve the issue.The procedure was completed with another of the same device.There were no patient complications.It was further reported that the lesion was located in the mid-superficial femoral artery and was 100% occluded with mixed morphology.
|
|
Manufacturer Narrative
|
Device eval by manufacturer: upon receipt at our post market quality assurance laboratory, this jetstream xc catheter was visually examined for any shaft damage and the functional testing of the device was completed.The device showed no damage.Functional analysis was done by completing the setup procedure per the instructions for use.Aspiration testing of the device was done per the test procedure.Test results showed that this device did not perform as designed per the test procedure specification sheet.Inspection of the remainder of the device revealed no damage or irregularities.The complaint was confirmed for evacuation failure.
|
|
Event Description
|
It was reported that aspiration was lost.This 2.4mm jetstream xc catheter was selected for use in a lower extremity angiogram procedure for peripheral artery disease.After approximately four minutes of run time, after three passes in blades down mode, and during the fourth pass in blades up mode halfway through the occlusion, aspiration appeared to be lost as if the catheter was clogged.Device suction could be heard; however, no bodily fluid or saline was being returned.Troubleshooting was performed but was unable to resolve the issue.The procedure was completed with another of the same device.There were no patient complications.It was further reported that the lesion was located in the mid-superficial femoral artery and was 100% occluded with mixed morphology.
|
|
Search Alerts/Recalls
|
|
|