BOSTON SCIENTIFIC CORPORATION INJECTION GOLD PROBE; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
Model Number M00560150 |
Device Problems
Failure to Deliver Energy (1211); Defective Device (2588); Detachment of Device or Device Component (2907); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/01/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Blocks d4, h4: the complainant was unable to provide the lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: imdrf device code a0501 captures the reportable event of distal tip detached.Imdrf impact code f2301 additional device required captures the retrieval of the detached piece.
|
|
Event Description
|
It was reported to boston scientific corporation that an injection gold probe device was used in the duodenum at the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) performed on (b)(6) 2023.After completing the ercp, the physician elected to use a gold probe for the slight bleeding at the ampulla site.The device stopped delivering energy while in use, and the entire tip detached during removal and remained inside the patient.The procedure was completed using another of the same device and the detached tip was retrieved with rat tooth forceps.There were no patient complications reported as a result of this event.Note: a photo of the device was provided showing the distal tip of the device detached.Additionally, the tip of the device and the working length can be seen kinked.
|
|