|
Model Number 1036 |
Device Problems
Degraded (1153); Low impedance (2285)
|
Patient Problem
Shock from Patient Lead(s) (3162)
|
Event Date 07/11/2023 |
Event Type
Injury
|
Event Description
|
A barostim system was implanted on (b)(6) 2023.On (b)(6) 2023, the patient presented to the emergency department experiencing shocking sensations near the neck incision.They stated they had experienced the shocking sensations a few times in the past, but it had occurred on and off all day on (b)(6) 2023.System interrogation showed a low impedance of 164 ohms, and the sensation was recreated during each impedance check or change to therapy settings.The root cause of the low impedance was unable to be determined, and the patient did not recall any falls, accidents, physical or massage therapy, or chiropractic visits.Therapy was disabled, and the patient was released without admittance.A csl replacement occurred on (b)(6) 2023.It was observed that the insulation was stripped from approximately 1/3 of the csl.The root cause of the lead insulation damage was unable to be determined.The patient reported they did not twiddle the device.As of (b)(6) 2023, it was reported the patient was no longer experiencing any adverse event.
|
|
Manufacturer Narrative
|
The reported csl was received at cvrx for analysis.The csl was received in three pieces.Two to three nicks were observed, which appeared to be from a blade, which may have occurred during explant.The distal section returned appeared to have been cleanly cut, such as by a blade, and did not have the appearance of a tear from twiddling or other manual manipulation.The distal section internal coil was stretched beyond the silicone tubing, with the silicone tubing being the length expected of a csl, indicating no part of the silicone tubing was missing.Functional testing of returned device is not possible due to the lead being returned in multiple pieces.Based on the information received, the cause of the reported event could not be conclusively determined.The device history and sterilization record for this device serial number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Cvrx id# (b)(4).
|
|
Search Alerts/Recalls
|
|
|