Model Number 0210 |
Device Problem
Connection Problem (2900)
|
Patient Problem
Unspecified Respiratory Problem (4464)
|
Event Date 07/26/2023 |
Event Type
malfunction
|
Event Description
|
The oxygen tubing has popped off the ventilator oxygen port several times when doing in line treatments on the circuits.
|
|
Manufacturer Narrative
|
The oxygen tubing on nebulizer popped off the ventilator oxygen port several times when doing in line treatments on the circuits.This would delay treatment for the patient.The criteria for reporting an adverse event have been met based on the reported information.
|
|
Manufacturer Narrative
|
The oxygen tubing on nebulizer popped off the ventilator oxygen port several times when doing in line treatments on the circuits.This would delay treatment for the patient.The criteria for reporting an adverse event have been met based on the reported information.The complaint of "oxygen tube disconnects from port during use" regarding part 210 was confirmed.The root cause was not determined but could possibly be a result of mating dimension incompatibility.A risk assessment was performed, and the ultimate risk was determined to be medium which would require reporting to the capa review board but ncmr - 03535 has been previously opened for this issue.There have been 4other complaints regarding the same part and a similar issue within the 24 months preceding the reporting of this issue.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
|
|
Event Description
|
The oxygen tubing has popped off the ventilator oxygen port several times when doing in line treatments on the circuits.
|
|
Search Alerts/Recalls
|