• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED Back to Search Results
Model Number N/A
Device Problems Calibration Problem (2890); Naturally Worn (2988); Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the motor speed was unstable, and per the images, there was some corrosion on the motor.Additionally, some screws and the reciprocating arm were worn, and the unit was out of calibration at the 0 reading.The motor, screws, and reciprocating arm were replaced, and the unit was recalibrated to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2: foreign: poland.
 
Event Description
It was reported during preventative maintenance, the device had a worn reciprocation arm, a calibration issue, worn screws and a motor speed issue which was explained as when the button was pressed, nothing happens.During evaluation, it was discovered that the motor speed was unstable.There was no patient involvement.Due diligence is complete there is no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME HANDPIECE, ELECTRICALLY-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17638642
MDR Text Key322182463
Report Number0001526350-2023-00995
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number30648800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2023
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/1995
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-