• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR 2 CANNULA; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR 2 CANNULA; CAT Back to Search Results
Model Number OJR416VT
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We have requested the return of the complaint optiflow junior 2 nasal cannula for investigation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in the netherlands reported, via a fisher & paykel healthcare representative (f&p), that one of the prongs of an optiflow junior 2 nasal cannula had broken off.There was no reported patient consequences.
 
Manufacturer Narrative
(b)(4).Method: the complaint ojr416 optiflow junior 2 nasal cannula was received at fisher & paykel healthcare (f&p) new zealand, where it was visually inspected.Results: visual inspection of the complaint cannula confirmed that the prongs were damaged.Conclusion: we are unable to determine the cause of the reported event.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: - appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death.- do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).- ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.- ensure the patient does not lie on the tubing as this may apply pressure to the patient's ears or face.- failure to apply and use this product within the directions, transport, storage and operating conditions specified in the labeling and user instructions may impair performance of this product or compromise safety (including potentially causing serious patient harm).
 
Event Description
A healthcare facility in the netherlands reported, via a fisher & paykel healthcare representative (f&p), that one of the prongs of an optiflow junior 2 nasal cannula had broken off.There was no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR 2 CANNULA
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17639690
MDR Text Key322148562
Report Number9611451-2023-00750
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOJR416VT
Device Catalogue NumberOJR416VT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/25/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-