• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P MCP PREFLEX IMPLANT SIZE 40 (STERILE PACKED); PROSTHESIS, FINGER, CONSTRAINED, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P MCP PREFLEX IMPLANT SIZE 40 (STERILE PACKED); PROSTHESIS, FINGER, CONSTRAINED, POLYMER Back to Search Results
Catalog Number MCPX-40
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Implant Pain (4561)
Event Date 07/31/2023
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery due to dislocating finger.
 
Manufacturer Narrative
The reason for this revision surgery, the agent reported "(dislocating finger)." the previous surgery and the surgery detailed in this event occurred 1 year 4 months apart.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocated finger.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCP PREFLEX IMPLANT SIZE 40 (STERILE PACKED)
Type of Device
PROSTHESIS, FINGER, CONSTRAINED, POLYMER
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key17647421
MDR Text Key322242345
Report Number1644408-2023-01162
Device Sequence Number1
Product Code KYJ
UDI-Device Identifier00886385025377
UDI-Public00886385025377
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberMCPX-40
Device Lot Number14608G
Was Device Available for Evaluation? No
Date Manufacturer Received10/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age63 YR
Patient SexMale
-
-