This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported ¿endoscope reprocessed in oer-6 automated endoscope reprocessor whose water filter had not been replaced within a recommended time¿ issue could not be determined, however, the issue is believed to be the result of improper user handling/user error as the replacement frequency in the subject facility was reported to be approximately every half a year, by contrast to every 2 months of replacement frequency which was recommended in the instruction manual,.The event can be prevented by following the instructions for use section below: oer-6 instructions, operation manual ¿chapter 7 section 3 replacing the water filter (maj-2317)¿.Olympus will continue to monitor field performance for this device.
|