• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CORONARY SAPHENOUS VEIN BYPASS GRAFT FOR EMBOLIC PROTECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CORONARY SAPHENOUS VEIN BYPASS GRAFT FOR EMBOLIC PROTECTION Back to Search Results
Model Number 44150
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2022
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: the wire was returned inside the delivery sheath.Visual and microscopic inspection observed that the spring tip was bent and detached.The filter bag came back in an undeployed state.No more damages were observed.
 
Event Description
Reportable based on device analysis completed on 07aug2023.It was reported that crossing difficulties were encountered.The 85% stenosed target lesion was located in the severely tortuous and mildly calcified lesion.A 190 cm filterwire ez was selected for use.During the procedure, the device was not able to cross the lesion.The device was removed and the procedure was completed with a different device.No complications were reported and the patient was stable.However, device analysis revealed tip detachment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERWIRE EZ
Type of Device
TEMPORARY CORONARY SAPHENOUS VEIN BYPASS GRAFT FOR EMBOLIC PROTECTION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17652487
MDR Text Key322436479
Report Number2124215-2023-45098
Device Sequence Number1
Product Code NFA
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/11/2024
Device Model Number44150
Device Catalogue Number44150
Device Lot Number0028855540
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
-
-