C.R. BARD, INC. (COVINGTON) -1018233 NITINOL CORE GUIDEWIRE, .035", 3 CM, STRAIGHT, STIFF, 150 CM LENGTH; NITINOL GUIDEWIRE
|
Back to Search Results |
|
Catalog Number 150NSS35 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/31/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
|
Event Description
|
It was reported that the nitinol guidewire was disassembled and found that the head end was bifurcated and could not be used.
|
|
Event Description
|
It was reported that the nitinol guidewire was disassembled and found that the head end of the was bifurcated and could not be used.
|
|
Manufacturer Narrative
|
The reported event is inconclusive as no sample was returned for evaluation.Although an exact root cause could not be determined a potential root cause could be inadequate material.A dhr review did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and state the following: "warning: inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract.Do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract.Do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
|