• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 NITINOL CORE GUIDEWIRE, .035", 3 CM, STRAIGHT, STIFF, 150 CM LENGTH; NITINOL GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 NITINOL CORE GUIDEWIRE, .035", 3 CM, STRAIGHT, STIFF, 150 CM LENGTH; NITINOL GUIDEWIRE Back to Search Results
Catalog Number 150NSS35
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/31/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Event Description
It was reported that the nitinol guidewire was disassembled and found that the head end was bifurcated and could not be used.
 
Event Description
It was reported that the nitinol guidewire was disassembled and found that the head end of the was bifurcated and could not be used.
 
Manufacturer Narrative
The reported event is inconclusive as no sample was returned for evaluation.Although an exact root cause could not be determined a potential root cause could be inadequate material.A dhr review did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and state the following: "warning: inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract.Do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract.Do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITINOL CORE GUIDEWIRE, .035", 3 CM, STRAIGHT, STIFF, 150 CM LENGTH
Type of Device
NITINOL GUIDEWIRE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17660666
MDR Text Key322372538
Report Number1018233-2023-06449
Device Sequence Number1
Product Code EZB
UDI-Device Identifier00801741051524
UDI-Public(01)00801741051524
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number150NSS35
Device Lot NumberNGHN2245
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-