Catalog Number SCCS1002
Device Problem Device Sensing Problem (2917)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/11/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the biomed stated the sensica device did not reading the bag during 3 step check and it would not recognized bag.The load cell was failed.
 
Event Description
It was reported that the biomed stated the sensica device did not reading the bag during 3 step check and it would not recognized bag.The load cell was failed.
 
Manufacturer Narrative
The reported issue was unconfirmed.The root cause of the reported issue could not be determined.The reported issue was not replicated in depot repair.The issue was not confirmed, updates were made as a part of the service procedure.The sensica device was evaluated.The issue was not replicated in the depot service.The sensica device underwent service updates, new ground wire and optical sensor gasket foam was applied.Load cell was failed.Reported issue was not confirmed.Service updates were made per service procedure.The device worked correctly out of the box as well as after the load cell recalibration/final functional verification.The reported event is unconfirmed, dhr review is not required.The reported event is unconfirmed, labeling/packaging review is not required.Correction: d,h.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA UO SYSTEM
Type of Device
SENSICA URINE OUTPUT SYSTEM
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
suite 200
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17661014
MDR Text Key322383264
Report Number1018233-2023-06455
Device Sequence Number1
Product Code EXS
UDI-Device Identifier00801741215773
UDI-Public00801741215773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSCCS1002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-