• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/58
Device Problems Device Dislodged or Dislocated (2923); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/14/2023
Event Type  malfunction  
Event Description
It was reported that during implant procedure, the novel right ventricular lead failed to be fixed and the lead became dislodged.It was alleged that the lead's helix became skewed.This lead was not installed during the procedure on (b)(6) 2023.There were no patient consequences.
 
Manufacturer Narrative
The reported events were unable to be fixed, lead dislodged and skewed helix.As received, a complete lead was returned in one piece with the helix found extended and clogged with dried blood/tissue.The reported event of skewed helix was not confirmed.Final analysis found over torque of the inner coil at the connector region consistent with procedural damage.No further issues were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17662933
MDR Text Key322417275
Report Number2017865-2023-40763
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502894
UDI-Public05414734502894
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/58
Device Lot NumberP000170176
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-