GIVEN IMAGING LOS ANGELES LLC MANOSCAN; SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) Back to Search Results
Model Number 1286
Device Problem Difficult to Remove (1528)
Patient Problem Sore Throat (2396)
Event Date 08/09/2023
Event Type  Injury  
Event Description
According to the reporter, the catheter was inserted via nostril without any resistance to 52 cm.The procedure went well with routine water swallow without any complications.When the catheter was pulled out, it got stuck at 46cm and would not retrieve further.At 46cm mark, a small tear on the plastic sheath was visible.To investigate further the patient was swiftly moved to the endoscopy room and anesthetized under the anesthetic with an endotracheal tube inserted.The endoscope went into the esophagus and the tip of the manometry catheter was sitting above the cardio esophageal junction as expected.No mechanical obstruction was seen in the entire esophagus or in the oropharynx.The white part of the catheter was not visualized, and the "obstruction" was felt to be in the nasal passage.The nasal passage was examined using a flexible nasal endoscopy.The patient has a deviated septum to the right nostril and enlarged turbinate.The torn plastic sheath was felt to be caught in the narrow passage preventing the catheter from being removed.The physician gradually maneuvered the catheter under visualization and managed to retrieve it with the sheath intact.No forceps were required.There was no immediate complication and the patient recovered well post-operatively apart from nasal congestion and sore throat.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the silicon sleeve had damage at the 46 cm mark.During the analysis, sensor 2 was failed on pico-scope test.Preliminary atp testing revealed that the catheter failed to calibrate on sensors 2 ,8,14 and 20.This failure every 6 sensors implies wire harness damage.It was reported that the catheter was difficult to remove from patient.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MANOSCAN
Type of Device
SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)
Manufacturer (Section D)
GIVEN IMAGING LOS ANGELES LLC
5860 uplander way
culver city CA 90230
Manufacturer (Section G)
GIVEN IMAGING LOS ANGELES LLC
5860 uplander way
culver city CA 90230
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17667489
MDR Text Key322473495
Report Number3005344223-2023-00003
Device Sequence Number1
Product Code FFX
UDI-Device Identifier04260167482484
UDI-Public04260167482484
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K151086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1286
Device Catalogue Number1286
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
Patient Weight45 KG
Patient RaceWhite
-
-