The test strips, meters a and b were received for investigation.The test strip vial, desiccant, and vial cap were inspected and were acceptable in appearance.There was no obvious reagent discoloration.Meter a and the returned test strips were tested using retention controls.Test results: level 1 ¿ 44, 44, and 44 mg/dl (qc range 30-60 mg/dl).Level 2 ¿ 308, 305, and 307 mg/dl (qc range 261-353 mg/dl).Meter b was tested using retention strips and retention controls.Test results: level 1 ¿ 43, 45, and 45 mg/dl (qc range 30-60 mg/dl).Level 2 ¿ 298, 294, and 299 mg/dl (qc range 261-353 mg/dl).All returned results were within acceptable range.The log files of both meters did not indicate any hardware or software issues that would lead to a measurement discrepancy.Meters a and b were disassembled for further investigation.There were no damages or contamination observed.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation is ongoing.
|