BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 2C8537 |
Device Problems
Fluid/Blood Leak (1250); No Flow (2991); Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that two (2) clearlink system continu-flo solution sets would not allow solution to flow from the catheter through the hub; the sets leaked from the hub site.The tubing would not connect to hub of the intravenous (iv) catheter despite changing the tubing out; it continued to leak.This was discovered during patient use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
E1: initial reporter address: (b)(6).G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
One device was received for evaluation; the other device was not received and therefore, could not be evaluated.Visual inspection using the naked eye did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.Functional tests were performed which included clear passage, pressure tests and the results were satisfactory.Additional tests were performed including priming and simulated use tests and no defects were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|