Model Number 11500A
Device Problems Gradient Increase (1270); Insufficient Information (3190)
Patient Problems Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Event Date 08/10/2023
Event Type  Injury  
Manufacturer Narrative
H10: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a patient with a 21mm 11500a aortic valve is being evaluated for a valve in valve procedure after an implant duration of two years, four months due to unknown reason.The procedure has not scheduled yet.
 
Manufacturer Narrative
The device history record (dhr) could not be reviewed, as the device serial number was not provided.Engineering evaluation summary: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potentials known and unknown patient-related contributing factors.Per technical summary 33069, rev a, structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.The most likely cause is patient factors, including a history of cabg and cad.All pertinent information available to edwards lifesciences has been submitted.
 
Event Description
It was reported and learned through medical records that a patient with a 21mm 11500a aortic valve underwent a valve in valve procedure after an implant duration of two (2) years, four (4) months and 21 days due to severe stenosis.The patient presented with chronic diastolic heart failure.The tavr was completed with a 26mm 9755rsl transcatheter valve without complication and patient was stable in icu.
 
Manufacturer Narrative
The device history record (dhr) was not able to be reviewed as the device serial number was not provided.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.Engineering evaluation summary: attempts have been made to obtain missing information; however, to date, no response has been received.Since there is no information regarding an allegation of a device malfunction, an engineering evaluation is not required.Based on the information available, a definitive root cause cannot be conclusively determined.All pertinent information available to edwards lifesciences has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INSPIRIS RESILIA AORTIC VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key17673907
MDR Text Key322545065
Report Number2015691-2023-15682
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number11500A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age76 YR
-
-