Trackwise#: (b)(4).The device was returned to the factory for evaluation on 08/23/2023.Photographs were provided by the account.A photographic evaluation was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the device in the photograph.An investigation was conducted on 08/29/2023.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the power supply.An electrical evaluation was conducted.The returned power cord was plugged into the device and the device was switched on.A pre-cautery test was performed per the instruction for use (ifu) with a reference adapter, vasoview hemopro 2 device, and the returned power cord and power supply vh-3010 at level 3.0.The device did passed the pre-cautery test.It did produce visible steam and heat during ten (10) 3-second activations.A tone was audible from the power supply upon activation.Based on the returned condition of the device as well as the evaluation results, the reported failure "electrical problem" was not confirmed.The reported device is an oem device.The certificate of conformance was reviewed for the serial#: (b)(6).The vendor certifies that this device lot conforms to all applicable product specifications and there were no non-conformances identified for the manufacturing batch.
|