Catalog Number 07976887190 |
Device Problems
Low Test Results (2458); Non Reproducible Results (4029)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The analyzer serial number is (b)(6).The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable elecsys ft4 iii assay results for 1 patient sample on a cobas e 801 module.The initial ft4 result was 0.838 ng/dl.The repeat result was 1.04 ng/dl.
|
|
Manufacturer Narrative
|
The calibration and qc data provided was acceptable.Based on the calibration and qc data, a general reagent issue could be excluded.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
|