AIZU OLYMPUS CO., LTD. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-H190
Device Problems Contamination /Decontamination Problem (2895); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, switch box scratched, air/water-cylinder has no color, suction-cylinder has no color, due to a crack on angle rubber, insulation resistance value at distal end does not meet the standard value, distal end cover chipped, and connecting tube cut.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the evis exera iii gastrointestinal videoscope had clogged channel.Upon inspection and testing of the returned device, foreign material was found clogged in the scope connecting tube due to insufficient cleaning.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.There was no report of patient harm or user injury associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The foreign object could not be identified.It is likely foreign objects remained in the channel due to the user's failure to properly reprocess the scope.If handled according to the instructions for use (ifu) below, the user may be able to detect the event: -inspection of the air-feeding function.-inspection of the objective lens cleaning function.-inspection of the suction function.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17686318
MDR Text Key323047025
Report Number9610595-2023-12734
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170305290
UDI-Public04953170305290
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-H190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2023
Initial Date FDA Received09/05/2023
Supplement Dates Manufacturer Received09/29/2023
Supplement Dates FDA Received10/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-