• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CENTURION VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CENTURION VISION SYSTEM, HANDPIECE TIP; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750853
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2023
Event Type  malfunction  
Event Description
A nurse reported that product cannot be tightened tightly before surgery.The product was replaced with another one and the surgery was completed.There was no patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
One opened phacoemulsification tip in a wrench, in a blister, was received for the report.The phacoemulsification tip was visually inspected and found conforming.The threads were functionally checked with a thread gage and the threads were found to be conforming.There was wear on the back of the flange, thread area and nut corners consistent with threading on a handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The photo attached to the parent file was reviewed by the manufacturing site.The photo shows a blister pack with reported lot number.Reported issue cannot be confirmed from photo.The returned sample was found to be conforming for all visual and functional testing associated with the reported event; therefore, a phacoemulsification tip cannot be tightened tightly was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the phacoemulsification tip was manufactured to specifications.All phacoemulsification tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, HANDPIECE TIP
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17689420
MDR Text Key322709522
Report Number2523835-2023-00460
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657508532
UDI-Public00380657508532
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750853
Device Lot Number14WNFL
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM, ACCESSORY, HANDPIECE
-
-