• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F; Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation Back to Search Results
Model Number A-TCSE-DF
Device Problems Fracture (1260); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2023
Event Type  malfunction  
Event Description
This report is to advise of a short circuit and exposed wires noted during analysis.
 
Manufacturer Narrative
One bi-directional, curve d-f, tacticath sensor enabled contact force ablation catheter was received for evaluation.The catheter shaft had been kinked 10.66¿ proximal to the distal tip and the shaft material was torn at the location of the kink, exposing internal components.The deformable body thermocouple (tc2) displayed high and variable resistance values during electrical testing due to the tc2 wires fracturing at the location of the kinked shaft.Additionally, a short circuit was also detected between tc2 and the tip thermocouple (tct), consistent with the tc2 wire fractures.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of short circuit and exposed wires remain unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE D-F
Type of Device
Catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key17701992
MDR Text Key322879280
Report Number3008452825-2023-00383
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027641
UDI-Public05415067027641
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA-TCSE-DF
Device Lot Number9065217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-