As reported, during hernia repair surgery, ventralight st w/ echo ps was inserted through the 12 mm trocar, and it was noted that one of the blue connectors was missing.It was stated by the customer that it appeared to have detached during insertion into the trocar.The mesh connector was found on the field, outside of the patient.It was reported that nothing was left inside the patient's body and there was no reported patient injury.
|
As reported, ventralight st w/ echo ps blue mesh connector detached, and it was located outside of the patient body in the field.The most probable cause is the connector inadvertently detaching from the mesh when being inserted into the trocar as reported.However, based on the information provided, and without having the sample to evaluate, no conclusions can be made.Review of manufacturing records confirms product was manufactured to specification.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned.
|