This event has been recorded by zimmer biomet under (b)(4).Evaluation and investigation are in process.Once the investigation is complete, a supplemental report will be filed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, d9, g3, g6, h2, h3, h4, h6, h10.Review of the most recent repair record determined the control bar was loose, the calibration was out at the 0 setting, and various components were worn.The motor, swivel, lever, eccentric shaft, poppet assembly, 4 and 3 in width plates, plunger, screws, bearings, gasket, pin, and seals replaced.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|