MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESONATE X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G447
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fever (1858); Unspecified Infection (1930); Fluid Discharge (2686)
Event Date 08/23/2023
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.Reportedly, the patient presented to the hospital with fever, drainage, and redness at the pocket site.There were no additional adverse patient effects reported.This crt-d was explanted.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection.Reportedly, the patient presented to the hospital with fever, drainage, and redness at the pocket site.There were no additional adverse patient effects reported.This crt-d was explanted.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against this device was not confirmed.This device was analyzed, passed all the baseline tests and exhibited normal device functions.This supplemental is being filed to capture d9: returned to manufacturer date and investigation results.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17705974
MDR Text Key322925160
Report Number2124215-2023-48875
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/07/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number305401
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age57 YR
Patient SexMale
-
-